DVVA II 4000
Appendix A
400-42803-01 (Rev B) - July 2017
A-15
6000 Series
6000- An error occurred while loading the application control information. Data entry
cannot be performed. (Code n)
6001-
Error accessing the sample information file (file name). (Code n).
6002- File (file name) cannot be analyzed. It is currently being edited.
Cause:
The control information disk file required by this application could not
be found.
Action A:
The disk drive may have failed or be corrupted. Run diagnostics on the
disk drive.
Action B:
Contact your service representative.
Cause A:
You attempted to open a file that is already open, possibly minimized.
Action A:
View the minimized icons, locate and maximize the file.
Cause B:
A computer or network problem occurred.
Action B:
Check the performance of your computer devices or network.
Cause C:
A software error occurred when the file was accessed.
Action C:
Contact your service representative.
Cause:
You selected a file for analysis that is still open in an editing window.
This is an information-only message and displays when you click
No
to
message 6005.
Action:
Click
OK
to close the dialog. Then, either continue editing the file, save
and close it; or click
Browse
and choose a different file for your anal-
ysis.
Summary of Contents for DVVA II 4000
Page 1: ...DVVA II 4000 Dynamic Void Volume Analyzer Operator s Manual Rev B 400 42803 01 July 2017 ...
Page 4: ......
Page 10: ...Table of Contents DV VA II 4000 vi 400 42803 01 Rev B July 2017 ...
Page 16: ......
Page 32: ......
Page 116: ......
Page 136: ......
Page 160: ......
Page 180: ......
Page 212: ......
Page 214: ......
Page 240: ......
Page 246: ......
Page 248: ......