6030- Unable to read the calibration file (file name).
6031-
Unable to write the calibration file (file name).
6032-
File contains calibration data for a different unit (S/N: nnn).
6035-
Analysis cancelled because there is not enough data for a blank correction.
Decrease the Recording Interval, increase the Ending Pressure, and/or decrease
the Scan Rate.
Cause:
You selected an invalid calibration file or one that cannot be read.
Action:
Be sure the media containing the calibration file has no problems.
Cause:
An attempt to Save calibration data has failed due to possible media
problems.
Action A:
Confirm the media containing the calibration file has no problems.
Action B:
Choose an alternate media to Save the calibration data.
Cause:
You loaded a calibration file that was created for a different unit.
Action A:
Load a file that was created for the current unit.
Action B:
Contact your service representative for assistance in calibration.
Cause:
The combination of values you entered for the
Scan
Rate
,
Ending
Pres-
sure
, and
Recording
Interval
does not yield enough data for a stable
blank correction.
Action:
Access the Analysis Conditions dialog and do one or more of the fol-
lowing:
• Decrease the Recording Interval
• Increase the Ending Pressure
• Decrease the Scan Rate
• Ensure that Use Prior Blank is not selected
Summary of Contents for DVVA II 4000
Page 1: ...DVVA II 4000 Dynamic Void Volume Analyzer Operator s Manual Rev B 400 42803 01 July 2017 ...
Page 4: ......
Page 10: ...Table of Contents DV VA II 4000 vi 400 42803 01 Rev B July 2017 ...
Page 16: ......
Page 32: ......
Page 116: ......
Page 136: ......
Page 160: ......
Page 180: ......
Page 212: ......
Page 214: ......
Page 240: ......
Page 246: ......
Page 248: ......