4. A series of prompts display requiring operator response. The length of time a test will run is
also indicated.
5. A popup window indicates the test is complete. Click
OK
.
6. Click the
Instrument to Regulator Shut-off Valve -1
tab to display its report.
The
Leak Rate field
should read
Passed
indicating the gas line from the analyzer to the
regulator is clean and leak-free.
If the
Leak Rate field
indicates
Failed
, a leak is indicated. Check the connections from the
analyzer to the gas regulator valve. Tighten as necessary, then run the test again.
7. Click the
Instrument to Gas Bottle Shut-off Valve - 2
tab to display its report.
The
Leak Rate
field should read
Passed
indicating the gas line from the analyzer to the
regulator is clean and leak-free.
If the
Leak Rate
field indicates
Failed
, a leak is indicated. Check the connections from the
regulator shut-off valve to the gas cylinder shut-off valve. Tighten as necessary, then run the
test again.
Clean and Verify the Gas Line
Gemini VII Operator Manual
239-42828-01 (Rev H ) — May 2021
10 - 9
Summary of Contents for GEMINI VII
Page 1: ...OPERATOR MANUAL 239 42828 01 May 2021 Rev H GEMINI VII SURFACE AREA ANALYZER ...
Page 10: ......
Page 16: ......
Page 92: ......
Page 126: ...6 About Reports SAMPLE LOG REPORT 6 34 Gemini VII Operator Manual 239 42828 01 Rev H May 2021 ...
Page 127: ...T PLOT REPORT t Plot Report Gemini VII Operator Manual 239 42828 01 Rev H May 2021 6 35 ...
Page 128: ...6 About Reports VALIDATION REPORT 6 36 Gemini VII Operator Manual 239 42828 01 Rev H May 2021 ...
Page 234: ......
Page 242: ......
Page 244: ......
Page 247: ......
Page 248: ......