![background image](http://html1.mh-extra.com/html/micromeritics/sedigraph-iii-5120/sedigraph-iii-5120_installation-instructions-and-checklist_1787652022.webp)
Installing the MasterTech 052 (Optional)
SediGraph III 5120 Installation
18
Mar 2010
6. When the rinse operation is complete, select
Unit [n] > Drain and load
.
7. Click
Continue
.
8. Fill the mixing chamber with baseline solution when instructed.
9. When the drain and load operation is complete,
Unit [n] >Baseline
.
10. Deselect
Prepare to load baseline liquid,
then click
Next
.
11. Click
Close
when the baseline operation is complete.
Installing the MasterTech 052 (Optional)
The MasterTech is an optional device for use with the SediGraph. If the customer purchased a
MasterTech, install it before proceeding. Refer to “Appendix A: MasterTech Installation” on
page 38 for instructions.