1 O
VERVIEW
This document contains procedures for installing and verifying the operation of the Smart VacPrep
degasser. Use these installation instructions and checklists to ensure that installation procedures
are completed properly.
It is recommended that installation procedures be performed in the order presented in
this document.
P
REPARE FOR
I
NSTALLATION
Before installation, ensure the operator has reviewed thoroughly the analyzer
Pre-installation
Instructions and Checklist
. Ensure all required equipment and supplies are unpacked and the
required personnel are available.
When the equipment is received, unpack and inspect the contents of the shipping container(s).
Use the packing list to verify that all products, accessories, software (if applicable), and
documentation are received intact and in the correct quantity. The shipping container(s) and
contents should be inspected within a few days of receipt in the event damage or loss has
occurred. Sort through all packing material before declaring missing equipment or parts.
Micromeritics recommends saving all shipping containers until installation of the equip-
ment is complete. All shipping containers where equipment is to be declared as dam-
aged or lost must be examined by the claims investigator prior to completion of the
inspection report.
1 Overview
Smart VacPrep Installation Instructions and Checklist
067-42870-02 (Rev A) — June 2022
1 - 1