2 Installation
C
ONNECT
G
AS TO
A
NALYZER
Analyzer
Gauge should indicate
VacPrep
15 psig (103 kPag)
Regulator Pressure Settings
Exceeding the maximum recommended air pressure could cause personal injury or
damage the instrument.
Ensure the gas cylinder is closed before connecting the gas line to the analyzer.
Ensure the gas cylinder is closed before connecting the gas line to the instrument.
1. Loosen, then remove the plug from the gas port on the back of the degasser
2. Insert the gas line into the port and hand-tighten the connector nut. Use a 7/16 in. (11 mm)
wrench to tighten the nut until very snug.
3. Connect all the remaining gases on available ports.
R
EMOVE
S
TABILIZING
T
UBES
The degasser ships with stabilizing tubes in heating stations 1 and 6 to prevent the heater block
from shifting during shipment. Ensure these tubes are removed before powering on the degasser.
2 - 4
VacPrep Operator Manual
061-42802-01 (Rev B) — Apr 2021