A T
EMPERATURE
C
ONTROLLER
E
RROR
M
ESSAGES
The temperature controller performs self-diagnostics when the degasser is powered ON and while
operating. If an error occurs during self-diagnostics, one of the following error messages will be
displayed.
LLLL
Cause:
Temperature has fallen substantially below temperature scale range or the tem-
perature sensor has failed (short circuit).
Action:
Observe the display. The temperature will be displayed if it is increasing. If the tem-
perature does not increase, the message will remain displayed. Power off the
instrument and contact the appropriate service personnel.
uuuu
Cause:
Temperature has risen substantially above temperature scale range or the tem-
perature sensor has burned out (open circuit).
Action:
Observe the display. The temperature will be displayed if it is decreasing. If the tem-
perature does not decrease, the message will remain displayed. Power off the
instrument and contact the appropriate service personnel.
A Temperature Controller Error Messages
VacPrep Operator Manual
061-42802-01 (Rev B) — Apr 2021
A - 1