R
EAR
P
ANEL
C
ONNECTIONS
A. Gas tubing to instrument
B. Gas regulator shut-off valve
C. Low pressure gauge
D. High pressure gauge
E. Gas cylinder shut-off valve
F. Regulator connecter nut
G. Regulator control knob
H. Brass reducer fitting
1. Insert one end of the power cord into the input power connector at the rear of the degasser
and the other end into the appropriate power source.
2. Attach an appropriate regulator to the gas cylinder. Leave the gas cylinder shut-off valve
closed until instructed otherwise.
3. If the regulator has a 1/8 in. outlet, proceed to Step 4. If the regulator has a 1/4 in. outlet,
attach the reducer fitting to the outlet of the regulator shut-off valve.
4. Tighten the regulator shut-off valve nut.
Do not overtighten the fittings. Doing so can collapse the brass ferrule and cause a
leak.
5. Attach the copper tubing to the brass reducer fitting.
6. Attach the other end of the copper tubing to the fitting on the back of the degasser.
Rear Panel Connections
VacPrep Operator Manual
061-42802-01 (Rev B) — Apr 2021
2 - 9