4 Degas the Sample
F
LOWING
G
AS
P
REPARATION
Samples that contain a high level of moisture could benefit from a higher flow rate
than 5 psi regulator pressure when using the flowing gas method. A flow rate up to
50 cm
3
/min (approximately 16 psi regulator pressure) may be used to aid in purging
moisture from the sample when using flowing gas preparation only.
1. Power ON the degasser.
2. Set the temperature controller to the correct temperature.
3. Turn the VAC/GAS control knob on the degas station to the GAS position.
A. Cooling stations
B. Degasser stations
C. VAC/GAS control knobs
D. Vacuum gauge
E. Temperature controller
4. Wipe the gas delivery tube with a clean, lint-free cloth to remove any residue left from the
previous sample.
5. Remove the stopper from the sample tube and insert the gas delivery tube into the sample
tube.
Do not pierce the sample bed with the gas delivery tube. Doing so could clog the gas
delivery tube. To insert the tube, tilt the sample tube nearly horizontal and tap the tube
to open a passageway through the sample. Then tilt the sample tube back to a ver-
tical position and tap the tube to settle the sample bed around the gas delivery tube.
4 - 8
VacPrep Operator Manual
061-42802-01 (Rev B) — Apr 2021