Flexible tubing connects the valves to the gas manifold inside the degasser. If there is no gas flow
to a particular valve, the tubing could be disconnected. To check the tubing:
1. Power OFF the vacuum pump.
2. Power OFF the degasser.
3. Remove the top rear panel.
4. The tubing should be connected to the valve and to one of the spokes on the manifold. If it is
disconnected at either end, connect it by pressing firmly into place. If a spoke on the man-
ifold is loose, tighten it finger-tight plus 1/2 turn with a wrench. Do not overtighten because
the gasket may become forced out of place causing the fitting to leak.
A. Flexible valve inlet
tubing (one per valve)
B. Gas manifold
C. Gas port
D. Vacuum port
E. Vacuum gauge tube
5. Replace the top rear panel.
6. Power ON the vacuum pump.
Valve Inlet Tube
VacPrep Operator Manual
061-42802-01 (Rev B) — Apr 2021
5 - 11