Chapter 12 Maintenance
8. Set your current source to the value that you want to use for the upper reference point and
wait 3-5 seconds until the value displayed by the DeviceConfig has stabilised. Ensure that the
required value is entered for the upper reference point before you press the button to adopt
the upper reference point for the relevant universal input (see ""Calibration" tab " on page
100). You can also use the "Set all upper" button to set the upper reference points for both of
the universal inputs at the same time. However, each of the universal inputs requires its own
current source for this purpose. The current sources must all be set to the same value during
this process.
9. Press the "Save" button to transfer the calculated offset and slope values to the myDa-
talog2Ex (see ""Calibration" tab " on page 100).
10. The "Print" button is used to generate the calibration protocol that must always be created to
ensure traceability. After pressing the "Print" button, a dialogue window will appear in which
additional information can be added. The report is printed once this window is closed by
selecting "OK". The standard printer is always used for this process without any further que-
ries. A PDF writer can also be used.
Rev. 02
101
Summary of Contents for 305001
Page 2: ......
Page 8: ......
Page 9: ...Chapter 2 Declaration of conformity Chapter 2 Declaration of conformity Rev 02 9 ...
Page 10: ......
Page 11: ...Chapter 3 Ex certification Chapter 3 Ex certification Rev 02 11 ...
Page 12: ...12 Rev 02 ...
Page 13: ...Chapter 3 Ex certification Rev 02 13 ...
Page 14: ......
Page 26: ......
Page 38: ......
Page 64: ......
Page 84: ......
Page 106: ......
Page 108: ......
Page 126: ......