Charger with power supply unit
1
Plug-in power supply (included in the scope of
delivery of 300697)
4
PSU Charger (300697)
2
Button (reserved for extension)
5
Status LED of the PSU Charger
Possible states of the status LED:
off: no PSU inserted
Flashing green: PSU is being charged
Green: Charging completed
Flashing red 1x every 5 seconds: PSU does not
contain a rechargeable energy store
Flashing red 2x every 5 seconds: PSU damaged
Flashing red 3x every 5 seconds: Supply voltage
of the charger is too low
3
Power supply unit e.g. AP (300524)
The charging process starts as soon as the power supply unit is inserted in the charger. If the status LED on
the charger flashes red once every 5 seconds, the power supply unit inserted in the charger does not contain
a rechargeable energy store. If the status LED flashes red three times every 5 seconds, the supply voltage of
the charger is too low. In this case, check the cable connection between the plug-in power supply and the
PSU Charger and check whether the plug-in power supply is correctly connected to an electric socket. If the
status LED flashes green, a normal charging process is in progress. The charging process is complete when
the status LED turns green.
If the status LED flashes red twice every 5 seconds, the power supply unit is faulty. Possible reasons for this
include a broken cable, short circuit or defective cells. In this case, the used power supply unit must be
replaced with a new one.
250
Rev. 05
Summary of Contents for myDatalogEASY V3
Page 2: ......
Page 13: ...Chapter 2 Declaration of conformity Chapter 2 Declaration of conformity Rev 05 13 ...
Page 14: ......
Page 42: ......
Page 76: ......
Page 88: ......
Page 102: ......
Page 110: ......
Page 116: ......
Page 234: ......
Page 244: ......
Page 252: ......
Page 254: ......
Page 266: ......
Page 276: ......