Chapter 9 User interfaces
Universal inputs
(Counter modes)
Cnt.Day
Filter time Time in [ms] during which the signal must remain
constant to initiate a level change. Used to suppress
brief faults (debouncing).
Reset at
Reset time of the day counter
Cnt.Intervl. Filter time Time in [ms] during which the signal must remain
constant to initiate a level change. Used to suppress
brief faults (debouncing).
Decay
Temporal function in the measurement cycle
off
Decay deactivated
min
The minimum of the last x
measurement values is
recorded.
max
The maximum of the last x
measurement values is
recorded.
avg
The arithmetic mean of the last
x measurement values is
recorded.
med
The median of the last x
measurement values is
recorded.
rms
The root mean square of the
last x measurement values is
recorded.
Time
Decay period. Up to a maximum of 60 measurement
valuescan be used for the calculation (e.g. 1 channel:
60 measurement values; ... 4 channels: 15 measurement
values each). To calculate the number of measurement
values that are taken into account, refer to "Example to
clarify the record interval, measurement cycle and burst
interval in conjunction with the decay" in chapter "Basic
settings" on page 96.
Rev. 06
69
Summary of Contents for myDatalogEASYplus
Page 2: ......
Page 8: ...Chapter 18 Contact information 139 8 Rev 06 ...
Page 9: ...Chapter 2 Declaration of conformity Chapter 2 Declaration of conformity Rev 06 9 ...
Page 10: ......
Page 22: ......
Page 40: ......
Page 62: ......
Page 101: ...Chapter 9 User interfaces 9 2 2 4 GPRS SIM tariff Selected SIM tariff Rev 06 101 ...
Page 102: ......
Page 110: ......
Page 112: ......
Page 118: ......
Page 120: ......
Page 136: ......
Page 138: ......