8. Analyse the fill level values determined during setup mode. If the displayed values do not match
the ones of the reference measurement, the zero point will need to be adjusted. To do so,
calculate the difference between the fill level determined by the BLE Gauge sensor module and
the reference measurement and enter the difference in the input field for the "Trimming"
parameter (see "Measurement channels" on page 83). As the "Trimming" parameter is added to
the measurement value of the BLE Gauge sensor module, a negative prefix must be set for the
"Trimming" parameter if the determined fill level value is greater than the reference
measurement (see "Additional explanation on the zero point adjustment and installation height of
the pressure sensor" on page 84).
Display of the fill level determined during setup mode.
1
Tool tip that specifies the fill level at a certain
time and indicates the average value
2
Average of the data generated during setup
mode.
9. Activate setup mode again if zero point adjustment was required. During the initial connection
established during setup mode, the changed configuration, i.e. the new "Trimming" value, is first
transmitted to the myDatalogGauge 3G and then to the BLE Gauge sensor module.
10. Wait until setup mode is terminated and then compare the determined fill level values with the
reference measurement again. If the values are still not the same, complete the zero point
adjustment again.
Note:
The current value of the "Trimming" configuration parameter must now be taken
into consideration when calculating the difference between the determined fill level and
the reference measurement.
9.4.2 Using the Bluetooth connection (BLE) and the myDatanet server
You will need the DeviceConfig configuration program (minimum version 4.21.4 ) to commission the
myDatalogGauge 3G . It can be downloaded free of charge from the following website:
www.microtronics.com/deviceconfig
Detailed instructions on using the DeviceConfig configuration program are provided in the chapter
"DeviceConfig " on page 101.
Note:
It is recommended that the myDatalogGauge 3G is first placed into operation in the office before moving
the measurement system to the place of use. During this process, you should set a site for subsequent
operation on the myDatanet server (see "Creating the site" on page 131) and determine a site setting (see "Site
configuration" on page 79). Take the opportunity to get to know the functions of the device in a stable
environment.
68
Rev. 01
Summary of Contents for myDatalogGauge 3G
Page 2: ......
Page 10: ...2 2 BLE Gauge 10 Rev 01 ...
Page 11: ...Chapter 2 Declaration of conformity Rev 01 11 ...
Page 12: ......
Page 13: ...Chapter 3 Ex certification Chapter 3 Ex certification 3 1 BLE 3G Gateway ATEX Rev 01 13 ...
Page 14: ...14 Rev 01 ...
Page 15: ...Chapter 3 Ex certification Rev 01 15 ...
Page 16: ...16 Rev 01 ...
Page 17: ...Chapter 3 Ex certification 3 2 BLE Gauge Rev 01 17 ...
Page 18: ...18 Rev 01 ...
Page 19: ...Chapter 3 Ex certification Rev 01 19 ...
Page 20: ...20 Rev 01 ...
Page 134: ......
Page 140: ......
Page 152: ......
Page 154: ......