Chapter 9 Initial Start-Up
17. Trigger the synchronisation between the myDatalogGauge 3G , DeviceConfig and myDatanet server
again to transmit the changed configuration, i.e. the new "Trimming" value, to the myDatalogGauge
3G and subsequently to the BLE Gauge sensor module.
18. If a zero point adjustment is required, repeat all of the steps from step 9 "Wait until some measurement
values have been recorded".
19. Then compare the calculated average value of the fill level with the reference measurement again. If
the values are still not the same, complete the zero point adjustment again.
Note:
The current value of the "Trimming" configuration parameter must now be taken into
consideration when calculating the difference between the determined fill level and the
reference measurement.
9.5 Testing communication with the device
9.5.1 Testing communication between the myDatalogGauge 3G and the myDatanet
server (mobile connection)
1. Within the desired customer, create a new site/application based on the IoT application
"myDatalogGauge" for operation of the myDatalogGauge 3G on the myDatanet server (see "Creating
the site" on page 131).
2. Configure the created site/application according to your requirements (see "Site configuration" on page
79).
The following settings should be carried out if you also want to test the communication with the BLE
Gauge sensor module at the same time:
1. Activate the required measurement channels by selecting the mode and specifying the serial
number of the BLE Gauge (see "Measurement channels" on page 83).
2. Check or enter the installation height of the pressure sensor in the input field for the "Sensor
Offset" parameter (see "Measurement channels" on page 83).
3. Check or enter the value for the zero point adjustment in the input field for the "Trimming"
parameter (see "Measurement channels" on page 83).
3. Connect the antenna (see "Connecting the GSM antenna" on page 62). The antenna is not included in
the scope of delivery and must be ordered separately.
4. Insert the activation module in the sensor recess of the myDatalogGauge 3G (see "Inserting the
activation module" on page 61).
5. To establish a connection, activate setup mode (see "Setup mode" on page 41) using the reed switch.
6. Wait until it is indicated in the list of sites/applications that the device is in setup mode. This is indicated
by a speech bubble with the "Setup" inscription.
Rev. 01
71
Summary of Contents for myDatalogGauge 3G
Page 2: ......
Page 10: ...2 2 BLE Gauge 10 Rev 01 ...
Page 11: ...Chapter 2 Declaration of conformity Rev 01 11 ...
Page 12: ......
Page 13: ...Chapter 3 Ex certification Chapter 3 Ex certification 3 1 BLE 3G Gateway ATEX Rev 01 13 ...
Page 14: ...14 Rev 01 ...
Page 15: ...Chapter 3 Ex certification Rev 01 15 ...
Page 16: ...16 Rev 01 ...
Page 17: ...Chapter 3 Ex certification 3 2 BLE Gauge Rev 01 17 ...
Page 18: ...18 Rev 01 ...
Page 19: ...Chapter 3 Ex certification Rev 01 19 ...
Page 20: ...20 Rev 01 ...
Page 134: ......
Page 140: ......
Page 152: ......
Page 154: ......