Chapter 9 User interfaces
9.2.1.6.2 Alarms
Warning
Value low
A warning is triggered, if the measurement value drops to or below this
value.
Value high
A warning is triggered, if the measurement value meets or exceeds this
value.
Alarm
Value low
An alarm is triggered, if the measurement value drops to or below this
value.
Value high
An alarm is triggered, if the measurement value meets or exceeds this
value.
Hyst %
Hysteresis for the all-clear in the event of an alarm/warning (e.g. Hyst=5%, alarm or
warning at 100 -> all-clear at 95) or reset the trigger (e.g. Hyst=5%, level = greater or
equal, trigger at 100 -> reset at 95)
9.2.1.6.3 Trigger
QU
Fastrecording (record interval = record interval / factor)
SL
Slow recording (record interval = record interval * factor)
RO
Switch on recording
RF
Switch off recording
XM
Initiate transmission
ON
Activate online mode
Level
Levels for initiating the trigger
Greater or
equal
The measurement value must be greater or equal to the level to
initiate the trigger.
Less or equal
The measurement value must be less or equal to the level to initiate
the trigger.
Rev. 01
71