Chapter 9 User interfaces
Note:
Additional explanation regarding the installation height of the sensor
Assumption:
Measurement range of the 4-20mA pressure sensor 0-5m. Please note that an mA sensor
requires the optional accessory "Pressure compensation tube with 1x 0/4-20mA 24VDC " (300255 ).
Installation situation of the pressure sensor
1
Installation height: 15 cm
3
Output value of the sensor: 5cm
2
Pressure sensor
4
Measured fill level: 20cm
Parameter
Value
Mode
Voltage
Type
4-20mA adapter
0%
0
100%
5
Sensor offset
0.15
Unit
m
Required configuration
Explanation:
As the "Sensor offs." parameter is added to the scaled measurement value, the installation
height that is used in the configuration provided in the example must be entered with a positive prefix.