4.4 Ex protection
The portable measurement instrument myDatasensMINI H2S 1000designed for use in areas with a
zone 2 explosive atmosphere.
The following conditions must be adhered to:
l
Only the manufacturer is permitted to open the housing to carry out maintenance work, which
must only be completed outside the Ex area.
l
Only the manufacturer's original spare parts must be used.
l
The battery must only be replaced with certified identical products.
l
Due to the electrostatic effects, the protective armour must not be rubbed with cloths in the Ex
zone.
II 3G Ex nA IIB T4 Gc
Important note: The Ex approval is only valid if the corresponding marking is provided on
the type plate of the measurement instrument .
Important note: The certificates of conformity and any relevant test certificates from the
respective authorities must be carefully observed during installation and commissioning.
1 myDatasensMINI H2S 1000
4.5 Safety instructions
For the connection, commissioning and operation of the myDatasensMINI H2S 1000, the following
information and higher legal regulations of the country (e.g. ÖVE), such as valid EX regulations as
well as the applicable safety and accident prevention regulations for the respective application case
must be observed.
Please read this manual completely before unpacking, setting up or operating this device. Observe
all hazard, danger and warning information. Non-observance can lead to serious injuries to the
operator and/or damage to the device.
14
Rev. 01
Summary of Contents for myDatasensMINI H2S 1000
Page 2: ......
Page 7: ...Chapter 1 Inhaltsverzeichnis_en Chapter 19 Contact information 107 Rev 01 7 ...
Page 8: ......
Page 9: ...Chapter 2 Declaration of conformity Chapter 2 Declaration of conformity Rev 01 9 ...
Page 10: ...2 1 Declaration of conformity Appendix 10 Rev 01 ...
Page 12: ......
Page 20: ......
Page 30: ......
Page 42: ......
Page 76: ......
Page 84: ......
Page 96: ......
Page 98: ......
Page 100: ......
Page 102: ......
Page 104: ......
Page 106: ......