Note: It is recommended that the myDatasensMINI H2S 1000is first placed into operation in the office before mov-
ing the device to the place of use. If you are considering transferring the data to a central myDatanet server in addi-
tion to processing data locally on your PC, you should create a site on the myDatanet server (see "myDatanet
Server Manual " 805002) and determine the site settings (see "Site configuration" on page 37). Take the oppor-
tunity to get to know the functions of the device in a stable environment.
1. The myDatasensMINI H2S 1000does not include a special transport mode. This means that the
device is always ready for operation and data is permanently recorded. Use the DeviceConfig
configuration program to adjust the record interval to your requirements (see ""Instrument"
tab" on page 52).
The following step is only necessary if you do not require the data generated up to the current point
in time.
2. Delete the data generated up to the current point in time by pressing the "Delete Meas-
urement Data" button that is located on the "Data" tab of the DeviceConfig configuration pro-
gram (see ""Data" tab" on page 60).
Important note: By clicking on the "Delete Measurement Data" button, all of the data
in the device memory and the measurement data assigned to the device on the PC are
deleted.
The following steps are only required if you want data to be transferred to the myDatanet server.
3. Create a site for operation on the myDatanet server (see "Creating the site" on page 80).
4. Configure the created site according to your requirements (see "Site configuration" on page
37).
5. Activate the synchronisation between myDatasensMINI H2S 1000, DeviceConfig and the myDa-
tanet server. For this purpose, use the button illustrated below that is located in the "Sync"
tab of the DeviceConfig configuration program (see ""Sync" tab" on page 58). The meas-
urement instrument is created in the pool of the myDatanet server during the first syn-
chronisation process (see "myDatanet Server Manual " 805002).
6. Link the myDatasensMINI H2S 1000with the created site (see "Site" on page 38 or "Creating
7. Initiate the synchronisation again so that the information detailing which site was linked to
the myDatasensMINI H2S 1000is transferred to the device.
32
Rev. 01
Summary of Contents for myDatasensMINI H2S 1000
Page 2: ......
Page 7: ...Chapter 1 Inhaltsverzeichnis_en Chapter 19 Contact information 107 Rev 01 7 ...
Page 8: ......
Page 9: ...Chapter 2 Declaration of conformity Chapter 2 Declaration of conformity Rev 01 9 ...
Page 10: ...2 1 Declaration of conformity Appendix 10 Rev 01 ...
Page 12: ......
Page 20: ......
Page 30: ......
Page 42: ......
Page 76: ......
Page 84: ......
Page 96: ......
Page 98: ......
Page 100: ......
Page 102: ......
Page 104: ......
Page 106: ......