"Calibration" tab
1 Lower reference point (target value in ppm)
7 Button for adopting the lower reference point
2 Upper reference point (target value in ppm)
8 Button for adopting the upper reference point
3 Raw value
9 Timer to determine the remaining waiting
time
4 H2S measurement value (calculated with the
currently valid values for the offset and
slope)
10 starts/stops the timer
1)
5 Calculated offset
11 Button to generate the calibration protocol
that should always be created to ensure
traceability.
6 Calculated slope
12 Button to transfer the values determined for
the offset and slope to myDatasensMINI H2S
1000
1)
After a restart, the timer is automatically reset to the value set via the two dropdown fields.
Rev. 01
55