1. Connect the USB radio transmitter (206.657) to the USB interface on your PC.
2. Start the DeviceConfig configuration program.
DeviceConfig
3. Connect the myDatasensMINI H2S 1000to the PC using the USB radio transmitter (206.657)
supplied (see "Establishing a connection to a device with a wireless M-bus interface" on page
49).
4. Additional tabs are displayed if the connection was established successfully. Now select the
"Calibration" tab.
myDatasensMINI H2S 1000specific tabs
5. Ensure that 0ppm is selected for the lower reference point. Wait five minutes before pressing
the button to set the lower reference point. You can use the timer integrated in the interface
to determine the amount of waiting time that remains. An acoustic signal is issued once the
timer has elapsed. During these five minutes, it must be ensured that the H2S concentration is
0ppm (if calibrated in a calibration room: extraction active and all of the valves on the gas cyl-
inders are closed; if calibrated with the Calibration set H2S 30ppm : environment free of H2S
gas).
88
Rev. 01