6. Place the myDatasensMINI H2S 1000on the calibration adapter, that is included in the scope
of delivery of the myDatasensMINI H2S 1000, and that should already be connected to the gas
extraction station (if calibrated in a calibration room) or the calibration gas cylinder (if cal-
ibrated with the Calibration set H2S 30ppm ).
Important note: Ensure that the myDatasensMINI H2S 1000is positioned as straight as
possible on the calibration adapter so that there is no gap between the protective
armour and calibration adapter.
Place the myDatasensMINI H2S 1000 in the calibration adapter
1 myDatasensMINI H2S 1000
3 Hose to the gas extraction station or cal-
ibration gas cylinder
2 Calibration adapter (included in the scope
of delivery of myDatasensMINI H2S 1000)
90
Rev. 01