9. The "Print" button is used to generate the calibration protocol that must always be created to
ensure traceability. After pressing the "Print" button, a dialogue window will appear in which
additional information can be added. The report is printed once this window is closed by
selecting "OK". The standard printer is always used for this purpose without any additional
prompt. A PDF writer can also be used.
Dialogue window to enter additional information
Customer
Name of the customer
Customer no. (ERP)
ERP number of the cus-
tomer
AB/RMA no.
Order number or RMA number that the device is
assigned to
Gauger
Name of the gauger who completed the calibration. The folder specified in the "Path for
signatures" must contain a PDF file with the digitised signature of the gauger. The file
name of the PDF must correspond exactly with the name of the gauger entered in the
"gauger" field.
92
Rev. 01