Chapter 9 User interfaces
9.2.1.3.1 Cross sensitivity of the H2S sensor
Gas
Concentration
Measurement value on the H2S sensor
Carbon monoxide
300ppm
6ppm
Hydrogen
10,000ppm
<15ppm
Sulphur dioxide
5ppm
<1ppm
Hydrogen cyanide
10ppm
-1.4ppm to -0.5ppm
Nitric oxide
35ppm
0ppm
Hydrochloric acid
5ppm
0ppm
Nitrogen dioxide
5ppm
-1ppm
Chlorine
1ppm
-0.05ppm to +0.04ppm
Ethylene
100ppm
0ppm
9.2.1.4 Internal channels
Int.
temperature
title
Freely selectable channel title for the internal device temperature
Unit
Selection of the temperature unit used by all of the server display
elements
9.2.1.5 Basic setting
Record interval
Time between measurement data recordings (read only)
The record interval can only be configured using the DeviceConfig
configuration program (see ""Instrument" tab" on page 52).
Time zone
Regional settings (not relevant for raw measurement data as this is stored
in the UTC)
Daylight saving time
Configuration for automatic time adjustment
Standard
The configuration for the time adjustment is adopted
by the global server setting (see "myDatanet Server
Manual " 805002).
Off
Automatic time adjustment deactivated
USA
Predefined setting for the American area
EU
Predefined setting for the European area
Default report
Selection of the report that is loaded by clicking on the device link in the
maps
Off
The default graphic is loaded.
"Name of a
report"
The selected report is loaded.
Rev. 01
39
Summary of Contents for myDatasensMINI H2S 1000
Page 2: ......
Page 7: ...Chapter 1 Inhaltsverzeichnis_en Chapter 19 Contact information 107 Rev 01 7 ...
Page 8: ......
Page 9: ...Chapter 2 Declaration of conformity Chapter 2 Declaration of conformity Rev 01 9 ...
Page 10: ...2 1 Declaration of conformity Appendix 10 Rev 01 ...
Page 12: ......
Page 20: ......
Page 30: ......
Page 42: ......
Page 76: ......
Page 84: ......
Page 96: ......
Page 98: ......
Page 100: ......
Page 102: ......
Page 104: ......
Page 106: ......