10.8 "Zero" tab
This tab enables you to compensate for offsets of the zero point. The zero point adjustment affects
the offset determined during calibration. The slope determined during the calibration remains
unchanged.
Detailed instructions on completing the zero point adjustment are provided in chapter "Zero point
adjustment" on page 94.
"Zero" tab
1 H2S concentration in ppm
In contrast to what is shown on the display
and measurement value graphics, negative
values are not cut off.
2 Button to complete the zero point adjustment
Rev. 01
53