6 |
Service part
PRODUCT FUNCTIONS
After inserting the batteries the welcome logo is displayed with a software version and the
equipment´s initialisation starts running. The first measuring and evaluation are performed.
Afterwards the equipment tries to make a radio connection with the switching unit. The time is
set on 12:00, date on 1.1.2009 and in case the connection is made, the first record with these
time details is registered. When performing backward monitoring it is possible to determine the
battery change realization.
The periodical recording of measured values and equipment status gets started. After making
the connection with the switching unit, the data are cyclically and in a set period transmitted into
the switching unit for recording. When the memory is full, the oldest data rewriting gets started.
It is not possible to delete the memory and it is either not possible to stop the recording – by this
the reliability of the equipment´s past status detection is ensured. Just before each record the
radio connection is made and the equipment status is detected.
After the period of measuring set in a ,,6´´ parameter has passed, the air humidity and -
temperature measuring is realized by an inbuilt sensor. Afterwards these data are evaluated. If it
is important to turn on/off the dehumidification, a radio signal is transmitted into the switching
unit.
In case of manual setting of the requested value the evaluation is also realized and if it is
important to intervene, the radio signal is sent.
In case of power cut of the switching unit, it is detected by the measuring unit only during the
next communication trial. After the power supply renewal, the switching unit´s status is activated
also only during the next communication trial. The time set in the ,,7´´ parameter (record period)
is the longest recovery time in case of constant operation. If the humidity or the requested
temperature gets changes, this time is shorter.
In case of power cut of the measuring unit, the switching unit detects its absence during the
parameter time 7 + 1 minutes. After this time the switching unit turns off the fan as well as the
compressor (if they were turned on before).
PAREMETERS´CONFIGURATION
For entering into the configuration menu, please keep the service button pressed and at the
same time press the control wheel. The service button is located under the display and is
accessible after removing the plastic cover.
For setting the requested parameter use the control wheel rotation.
Press the wheel for confirmation and continuing on the next configuration side.