en - Areas of application
44
Reprocessing
transmission
instruments
Check that internal sections of transmission instruments are free of
blockages.
Insert the transmission instrument into the holder. The instrument is
then held in position by the spring clamp.
After reprocessing pull the transmission instrument out of its holder.
The spring clamp will automatically spring back to its starting
position.
After reprocessing, the insides of the instruments must be dried
with sterile compressed air before being cleaned and sterilised as
appropriate in accordance with the manufacturer’s instructions.
Please observe national regulations.
Before using transmission instruments again following repro-
cessing, a function check must be carried out, e.g. by spraying into
a basin, to ensure they are clear.
AUF 1 holder for transmission instruments
AUF 1 holders are designed for reprocessing dental transmission in-
struments.
Optional ac-
cessories
– A 801, threaded adapter
– ADS 1, silicone adapter, white, for holding transmission instruments
with a
of approx. 20 mm
– ADS 2, silicone adapter, green, for holding transmission instruments
with a
of approx. 16 mm
– ADS 3, silicone adapter, red, for holding transmission instruments
with a
of approx. 22 mm
– FP 20, 20 filter plates, porosity 2,
30 mm