en - Areas of application
46
Replacing
filter plates
Filter plates must be replaced
approx. every 2 weeks
or after
20 reprocessing cycles
have been completed.
Unscrew the holder completely.
Dismantle the holder and dispose of the filter plate
.
Rinse out the lower part
thoroughly.
Insert a new filter plate.
Damaged or used filter plates must not be inserted under any
circumstances.
Screw the upper and lower parts of the holder back together and fit
the holder back onto the injector manifold.
A 814 holder for the tips of air scalers and piezo scalers
A 814 holder
A 814 holder with A 813 adapter
The A 814 holder enables the tips of air scalers and piezo scalers to
be reprocessed. To increase capacity, the holder can be combined
with the A 813 adapter. The holder and the adapter each come with
their own operating instructions.