en - Areas of application
35
Compressed air in-
dicator
The compressed air indicator on the pressure chamber shows
whether a sufficient amount of pressure has been provided.
If the red indicator of the compressed air indicator pokes out, the
pressure is sufficient.
X
If the indicator does not poke out, the pressure is not sufficient.
If the connection between the quick-release couplings on the pres-
sure chamber and the hoses for the air regulators is not be abso-
lutely tight, the pressure will drop over time.
If the pressure drops too much, the wash water may enter the air
regulators.
Therefore, supply the pressure chamber with pressure as close to
the programme start as possible.
Placing the insert in the load carrier
Lift the insert at the side handles.
Place the insert in the load carrier of the dishwasher or the ma-
chine.
After reprocessing
It is essential that you pay attention to the “Application technology”
chapter in the operating instructions for the dishwasher or the ma-
chine and the instructions specified by the manufacturer of the air
regulators.
Summary of Contents for APFD 200
Page 2: ...2 de 4 da 15 en 26 fi 37 fr 48 it 60 nl 71 no 82 sv 93 ...
Page 103: ......