Do not cut the inlet hose or
damaged in any way.
^
See also the installation diagram
supplied.
DI-Water connection (pressure
resistant) (over 9 to 145 psi)
The machine comes with a standard
connection for a pressurized DI system.
The recommended water pressure (flow
rate) is 25 - 60 psi, however, water
pressure of 9 - 147 psi is acceptable.
^
The DI pressure-tested hose (marked
"H
2
O pure") with a
3
/
4
" (2 cm) hose
thread must be connected to the
on-site water valve for DI purified
water.
,
If the machine will not be
connected to DI water, the cleaning
and disinfection programs must be
changed by Miele Technical Ser-
vice. The intake hose stays at the
rear of the machine.
DI water connection (pressureless)
1.25 - 9 psi (optional)
For connection to 1.25 - 9 psi water
pressure the machine must be
converted if it was not ordered from the
factory as a special model. A feed
pump must be installed by Miele's
Technical Service Dept.
The discharge nipple of the DI water
container (pressureless) must be at
least at the same height as the
machine's upper edge.
Plumbing