Plumbing
*INSTALLATION*
160
IMPORTANT
UK, Australia and New Zealand
For the UK, Australia and New Zealand a non-return check valve is
required between the tap and filter.
Do
NOT
shorten or otherwise damage the inlet hoses (see
diagram).
See installation diagram supplied.
Retrofitting the
large-surface
filter
If the water contains a high level of insoluble components, a large-
surface filter can be installed between the stopcock and the water
inlet hose.
The large-surface filter is available from the Miele Customer Service
Department.
AD water
connection for
30-1,000 kPa (UK:
100-1,000 kPa)
pressure -
pressure-
resistant
(optional)
This machine can be optionally supplied for a pressurised system
operating between 30-1,000 kPa (UK: 100-1,000 kPa). If the water
pressure is below 200 kPa the water intake duration will be
automatically increased.
The pressure tested hose for AD water, coded green, has a ¾ inch
threaded union for connection to the onsite stopcock for AD water.
If the machine is not going to be connected to demineralised
(DI) water, the DI water connection has to be deactivated by the
Miele Customer Service Department. The intake hose remains on
the back of the machine.
AD water
connection for
8.5-60 kPa -
without pressure
(optional)
The machine has to be converted for connection to 8.5-60 kPa
pressure unless ordered as such at the factory. Installation of the
pump must only be carried out by Miele Service.
With a pressureless AD water connection, the drainage point must
be at least as high as the top of the machine. See installation
instructions.