Please note! When a large quantity of demineralized water is
refilled, the floating body does not immediately indicate the
correct filling level as the specific density of demineralized
water is lower than that of the electrolyte. Immediately after
filling of a large quantity of demineralized water the floating
body indicates a filling level which is lower than the actual level.
Therefore, refilling of demineralized water must be done step by
step. The unit must be switched on and the gas production
started for approx. 30 to 60 seconds between each refilling
step.
1. If the floating body in the filling duct is no longer visible, refill
max. 0.2 litre demineralized water.
2. Switch on the unit at the
on/off
key
I/O
at the operating
panel to start the gas production. The filling duct can remain
open.
Caution! Risk of fire and explosion! Keep away from
ignition sources!
3. Switch off the gas production after approx. 30 seconds at
the
on/off
key
I/O
at the operating panel.
4. Check the filling level again:
a. If the floating body in the filling duct is still not visible,
repeat steps 1. to 3.
b. If the floating body in the filling duct is now visible refill
only a small quantity of demineralized water until the
floating body is level with the filling duct edge.
5. Screw the closing cap back onto the filling duct and tighten
it.
Illustration 8.1.1. Filling duct with glass floating body at maximum
filling level
NOTE
How to proceed:
refilling of
demineralized water
A
B