Guidance and manufacturer’s declaration - electromagnetic immunity
Conducted disturbances
induced by RF fields
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V
in ISM and amateur radio
bands between 0.15
MHz and 80 MHz Modu-
lation 80 % AM at 1 kHz
3 V
0.15 MHz to 80 MHz
6 V
in ISM and amateur radio
bands between 0.15
MHz and 80 MHz Modu-
lation 80 % AM at 1 kHz
As described in annex A.
3 of IEC 60601-1-2, the
voltage induced on a
cable in the frequency
range 150 kHz to 80
MHz from a field strength
of 10 V/m is unlikely to
exceed 6 V r.m.s. There-
fore, assuming a field
strength E of 10 V/m in
the ISM and amateur
radio bands between
0.15 MHz and 80 MHz
(see note 2), the min-
imum separation dis-
tance d in m to a RF
communications equip-
ment with a maximum
power P in W can be cal-
culated using the equa-
tion in clause 8.10 of IEC
60601-1-2:
d = 6 / E ∙ √P
Voltage dips
IEC 61000-4-11
0% U
T
; 0.5 cycle at 0°,
45°, 90°, 135°, 180°,
225°, 270° and 315°
0% U
T
; 1 cycle; 70% U
T
;
25/30 cycles; Single
phase: at 0°
0% U
T
; 0.5 cycle at 0°,
45°, 90°, 135°, 180°,
225°, 270° and 315°
0% U
T
; 1 cycle; 70% U
T
;
25/30 cycles; individual
phase: at 0°
Mains power quality
should be that of a typ-
ical commercial or hos-
pital environment. If the
use of the miha-bodytec
m.ove EMS training
device requires con-
tinued operation during
power mains interrup-
tions, an uninterruptible
power supply (UPS) suf-
ficiently fast and powerful
to provide the energy
needed shall be used.
Voltage interruptions
IEC 61000-4-11
0% U
T
; 250/300 cycles
0% U
T
; 250/300 cycles
Electromagnetic compatibility
2021-07-28
miha-bodytec m.ove EMS training device