TIR>TIS OR TCR<TCS cycle end after reaching preset time value, or after drop
of temperature in bar below preset value
TIR>TIS OR HUR<HUS cycle end after reaching preset time value, or after drop
of humidity below preset value
TIR>TIS AND
TCHR<TCH
cycle end after reaching preset time value and after drop
of temperature inside chamber below preset value
TIR>TIS AND
TCR<TCS
cycle end after reaching preset time value and after drop
of temperature in bar below preset value
TIR>TIS AND
HUR<HUS
cycle end after reaching preset time value and after drop
of humidity below preset value
INn=1
cycle end when “end release” is on
TIR>TIS AND INN=1
cycle end after reaching preset time value, and “end
release” must be on
TIR>TIS OR INN=1
cycle end after reaching preset time value, or after
switching on “end release”
INn=0
cycle end when “end release” is off
TIR>TIS AND INN=0
cycle end after reaching preset time value, and “end
release” must be off
TIR>TIS OR INN=0
cycle end after reaching preset time value, or after
switching off “end release”
-
press “Enter”
-
complete setting of parameters for one step by pressing “Stop” key
-
select next step to edit, or press “Stop”
and finish editing step parameters
5.2.3. Alarm setting
21 alarms my be activated in the controller:
-
11 from control inputs
-
5 from sensors
-
5 when measurements exceed allowable values
The following items may be defined for each alarm:
-
name
-
relay status
-
alarm delay time – time from alarm detection to its activation