BoneSTATION
MM093-002
– Operator Manual
16
2.3. Placement
BoneSTATION can be positioned in its final place of use. Make sure it is firmly placed and the bench is level.
During the setup, BoneSTATION must be disconnected from the main power supply.
A correct location of BoneSTATION is under a fume hood.
Provide a bench of at least 44cm (17.
3”) wide and 59cm (29”) deep and place BoneSTATION on it.
A proper location of BoneSTATION is under a fume hood.
2.3.1. Device overview
The picture below shows the front view of BoneSTATION system.
The device consists of:
•
A FRONT MODULE: a stainless-steel ceramic coated hot plate.
•
An INFRARED TEMPERATURE SENSOR installed behind the hot plate for non-contact temperature
measurement.
•
A BACK MODULE: a holding position with magnetic stirrer for a second histomodule, WITHOUT
HEATING.
•
A SAMPLE ZONE including two removable Petri glass containers. These items are all fully
washable.
•
An LCD TOUCH SCREEN 7
” user interface.
•
A WARNING LAMP to warn users when the plate is hot.
•
A Brushless magnetic stirrer.
FRONT
MODULE
HOT PLATE
BACK
MODULE
HOLDING POSITION
INFRARED
SENSOR
WARNING
LAMP
LCD touch
screen
SAMPLE
ZONE
GLASS
CONTAINER
GLASS
CONTAINER