2.4. Fumes extraction system
1.
Connect the black exhaust tube (CODE 50332) to the exhaust outlet located on the back panel of the
unit.
In case of connection to an external fume extraction system, the customer is required to adapt the hose
connector. The plastic hose supplied has these characteristics:
Length: 2.5m(98in)
External diameter: 55mm(2.15in)
Internal diameter: 50mm(2in)
Fan air flow range:23
–27m
3
/h (cube meter per hour).
2. Use the metal cable clamp (CODE 64757) to fix the hose to the exhaust outlet connection. Use a
screw driver to tighten the metal cable clamp.
3. The other end of the exhaust hose is normally connected to any vapors extractor system.
DO NOT connect the HistosMATE unit to the main power supply during the installation.
The external fume extraction system must be properly installed at all times, according to
local regulations on the safety of laboratory in force.
The external fume extraction system must permit an air flux 27m
3
/h (cube meter per hour).