Milestone s.r.l.
MM063-008
– HistosMATE – Operator Manual
14
2.5. Power supply
A ground connection is available in the lab to assure the unit works correctly (yellow/green, green or
bare wire). Do not connect a Milestone unit without ground connection. Do not use the blue/white
wire (neutral of power line).
Power supply neutral has to be referred to ground.
Power supply line wires (their size and distance from the switch board) used in the laboratory are
compliant with the label positioned on the back side of our units.
Plug/socket is visible during the normal use and easily accessible.
Before connecting the plug to the laboratory’s socket, check that the power supply
voltage of the unit matches the power supply line of the laboratory.
Damages occur when wrong power supplies the instrument.
The power supply must be protected by a residual current circuit breaker with over
current protection, residual current 30mA. Power line must have potential earth
connection.
CONNECTING CABLES: Never connect or remove cables when power to the unit is
switched on, this will avoid risk of damage.
Connect the power cord socket in the HistosMATE plug and the power cord plug to main power supply
socket.
2.6. Reagents
Complete the set up with reagent, by providing at least:
10L of cleaning reagent
Main plug is considered as disconnecting device
Equipment shall be connected to main socket outlet which is accessible and visible by the
operator.
Disconnect main plug from the socket outlet before setting up the system and before
cleaning.
The above reagents must be onsite and available at time installation of unit.
Summary of Contents for HistosMATE
Page 1: ...HistosMATE Automatic cleaner OperatorManual MM063 MILESTONE H E L P I N G P A T I E N T S ...
Page 2: ......
Page 25: ......
Page 26: ......
Page 27: ......