– HistosMATE – Operator Manual
17
3.2. Running a cleaning cycle
The following table shows the standard protocol for a cleaning cycle, to be carried out with a xylene
substitute made of Isoparaffins (for details see chapter 1.7).
REAGENT
CYCLE TEMPERATURE
CYCLE TIME
ISOPARAFFIN
60°C
7 min
Place the rack (or the stainless steel basket for moulds and forceps
– CODE 65237) on the relative holder
(CODE 65235) connected to the lid and place it into the warmed solution.
The holder is compatible also for the synergy racks.
Synergy rack for Logos (CODE 66810)
Synergy rack for Pathos Delta (CODE 66811)
If you have a previous model of HistoMATE (different to CODE 65220/SY or CODE
65201/SY) can be updated to the new simply by ordering the new model kit upgrades
synergy (CODE 66870) which includes stainless steel basket and relative holder see
below image.