1.4. Transportation and storage conditions
Temperature: -20°C to +60°C.
Humidity:
up to 80% (up to 50% at 40°C
– 104°F) non-condensing.
Before using HistosMATE (when coming from a storage room), let the unit reach working
environmental conditions (at least half an hour).
1.5. Warning information
Should the equipment be used in ways different than those specified by the manufacturer, the warranty
provided by the manufacturer may be void.
Main power plugs are considered disconnecting device. Disconnect all plugs from the socket outlet before
assembling the system, connection of the accessories and cleaning.
Equipment shall be connected to a main power socket outlet which is accessible and visible by the operator.
All parts of the equipment and all accessories must be supplied only by the manufacturer.
DO NOT OPEN THE CHASSIS. Failure to follow the instruction in the documentation
might lead to a reduction of device safety. If the equipment is not working properly,
please contact your supplier or the manufacturer.
Do not remove the instrument panels. Only authorized and qualified SERVICE
personnel may repair the instrument and access the instrument’s internal
components.
It is important that normal standards for safety and good laboratory practices are
applied. Always use common sense and the best known practice when operating the
instrument. Failure to follow the instructions in the documentation might lead to a
reduction of device safety.
The Competent Authority owning the instrument has primary responsibility for
accident-free operation together with designated personnel who operates SERVICE
or repair it.
To avoid damage to the instrument, use only the reagents listed in chapter 1.7
“Compatible Reagents”. Failure to follow the instructions in the documentation
might lead to a reduction of device safety.
Be careful: exhaust reagents are toxic.
Handle with care and follow the local standard regulations.
Use caution when handling hot, warm reagent.
Never install the HistosMATE unit under a fume hood. The presence of acids or
caustic reagents might contaminate the air flow used for ventilating the electronic
components, exposing them to possible corrosion or shorting of circuits.
Summary of Contents for HistosMATE
Page 1: ...HistosMATE Automatic cleaner OperatorManual MM063 MILESTONE H E L P I N G P A T I E N T S ...
Page 2: ......
Page 25: ......
Page 26: ......
Page 27: ......