Milestone s.r.l.
MM063-008
– HistosMATE – Operator Manual
8
1.6. Instrument labeling explanation
For symbols explanation see chapter 1.1.
1.7. Compatible reagents
Regenx® or Diasolv® or MicroClean® or Clear Rite3® or Soltrol® or other xylene substitutive solution,
(made of isoparaffins).
1.8. Waste disposal of the equipment
In case this system needs to be returned to the manufacturer for any repair, it is necessary to clean and
disinfect the unit before return. Non-disinfected devices will not be accepted and you will be contacted to pay
the relative disinfection cost.
International health rules require that shipments of biohazard materials are not done in standard packages
(risk of sanctions).
It is recommended to use enzymatic detergents and polyphenolic based disinfectants or chlorine substitutes
to clean and disinfect the instrument. Alternatively use of a solution of 1:10 Bleach (recommended by the
CDC for disinfecting) is also allowed to decontaminate the instrument.
Do not use XYLENE, acetone, benzene, or trichlorethane in the instrument. They can
seriously damage the unit.
DO NOT USE REAGENTS DIFFERENT FROM THOSE LISTED ABOVE.
In case of doubts please contact:
application@milestonemedsrl.com
.
It is necessary to put 10 L of cleaning reagent in the HistosMATE cavity.
Do not remove the side panels of the unit during cleaning operations.
Disconnect the power supply before any operation.
Do not use direct or high-pressure water to clean the instrument.
For safety reason, any operation without the individual protective devices must be
avoided.