ULTRASONIC PORTABLE MILK ANALYZER
www.milkotester.com
-13-
Operating Instructions Master Pro
The results will be printed immediately after they are ready if the
printer is in automatic mode. If the printer is in manual mode, pressing
the UP will cause printing results.
Changing modes is done by Menu ->Settings->Printer En/Dis
Page2
By choosing the ENTER button you can
start measuring anew.
By choosing the EXIT button you can start
from the very beginning and choose to
analyze another type of milk.
When finished analyzing the instrument issues a beep sound, returns
the analyzed sample material back into the cup, and the values of all
measured components appear on the display. If the unit is connected
to the printer all results will be automatically printed out.
At this point the sample cup can be removed from under the pipette.
The instrument will continue to show the measured values until a new
analysis is initiated.
WARNING!
Make sure that the instrument is at rest during analysis. Any shaking of
the device will cause inaccurate results.
NOTE:
Due to the possible presence of water inside the flow system left from
the last flushing procedure, we recommend that you do not take into
account the first analysis after flushing if they appear inaccurate. To
prevent or reduce this inaccuracy due to water in the system, please
refer to chapter “Cleaning and Maintenance”, section “Thorough
Cleaning”.
To access the menu, press the MENU button. The range of functions
is grouped into submenus. Scroll through the menu to select the one
that you need and then select the settings you need to make. Press
ENTER to activate them. Press EXIT to keep the previous settings.
4.3 MAKING CORRECTIONS AND CALIBRATION
In the process of work with the analyzer there is a possibility the
results to start differing between the data for some of the measuring
parameters when measured with the milk analyzer and the
Menu