M
AINTENANCE
- 59 -
I
NSTALLING THE NEW
UV
L
AMP
Ensure that you use the gloves supplied with the
UV replacement kit. Plug the electrical cable to
the new UV Lamp (G).
Carefully insert the UV Lamp into the UV
housing (H).
Attach the UV housing with the Velcro belt (I).
Install the SmartPak (J).
Install the Final Filter.
If the SmartPak is not being replaced, then
reinstall the old SmartPak BEFORE powering ON
the system. Otherwise, the system will go into
FLUSH mode for 15 minutes during which no
Product Water will be available.
Close the front cover.
Plug the electrical cord to power ON the system.
Reset the UV Timer. See
Section 6-10 How to
View or Reset the UV Lamp Timer (C05).
G
H
I
J
ATTENTION