PL-573
AiRanger SPL
Page 91
Technical Refer
e
nce
Parameter
Values Dependent on Measurement Response (P003) Value
#
Name (Units)
1
(slow)
2
(medium)
3
(fast)
4
(surge)
5
(imm.)
P070
Failsafe Timer
(min)
100
10
1
0.1
0
P700
Max Fill Rate
(m/min)
0.1
1
10
100
1000
P701
Max Empty Rate
(m/min)
0.1
1
10
100
1000
P702
Filling Indicator
(m/min)
0.01
0.1
1
10
100
P703
Emptying Indicator
(m/min)
0.01
0.1
1
10
100
P704
Rate Filter (option)
4
3
2
2
2
P710
Fuzz Filter (% of P007)
100
50
10
1
0
P711
Echo Lock (option)
1 or 2
(dependent on Material, P002)
0
0
P713
Echo Lock Window
(dependent on P701/P702 and time since last valid measurement)
P728
Shot Delay (s)
0.5
0.5
0.5
0.2
0.1
P803
Shot/Pulse Mode (option)
2
2
2
1
1
P841
Long Shot Number (qty)
10
5
2
1
1
Application Examples
The following examples describe the use of SPL programmable features, to meet the
needs of specific process measurement requirements.
It is highly unlikely that your specific installation will match one of these examples
exactly.
However, by reviewing the example (or combination of examples) that matches your
installation best, the relationship between SPL features and process measurement
requirements may be more easily recognized.
Example #
Description
1
Level (or Material Volume) Measurement for a single vessel.
2
Space (or Remaining Vessel Capacity) Measurement for a single
vessel.
Summary of Contents for AiRanger SPL PL-573
Page 1: ...AIRANGER SPL Instruction Manual PL 573 March 2001 IRANGER SPL 33455730 Rev 1 2 ...
Page 8: ...Page 8 AiRanger SPL PL 573 Specifications ...
Page 24: ...Page 24 AiRanger SPL PL 573 Installation ...
Page 86: ...Page 86 AiRanger SPL PL 573 Enhancement Param ...
Page 102: ...Page 102 AiRanger SPL PL 573 Maintenance ...
Page 108: ...Page 108 AiRanger SPL PL 573 Troubleshooting ...