The safe operation of the product is also related to the principles listed below:
The device may only be operated in sterilisation mode in a tightly closed room (no ozone may penetrate outside
the sterilised room). Therefore, make sure that all windows, doors and any other escape routes for ozone from the
room are tightly closed. Gaps around and under doors and ventilation grilles must be carefully sealed. The room
must be protected from access by third parties.
It is also forbidden to stay in rooms adjacent to the ozone room during the ozoning process. If the ozoned room is
not properly sealed, some ozone may penetrate into other rooms.
To protect the room from access by third parties, it is essential to place warning signs “WARNING! NO ENTRY!
STERILISATION IN PROGRESS". Place them in a clearly visible place. The signs are included in the package.
After starting the sterilisation mode, leave the room where the steriliser is operating as soon as possible. The time
to leave the room is 120 seconds from the start of the ozoning function, only after that time the ozone generators
start. It is forbidden to stay in the room during operation of the ozone generator without a full respiratory and eye
mask with a suitable absorber. It is recommended to use a protective mask according to PN-EN 136 with an
absorber according to PN-EN 14387, type NO P3. It is also recommended to use personal protective equipment
for the skin according to the table of personal protective equipment.
Smoking and the use of tools that cause a flame or spark is prohibited in an ozoned room.
In order to deactivate the sterilisation mode before the end of the process, one should enter the room wearing a
full respiratory and eye mask with a suitable absorber (in accordance with EN 136 and EN 14387 standards) and in
the skin protection equipment in accordance with the personal protective equipment table.
When the ozoning of the room is complete, it must be thoroughly ventilated for at least 30 minutes.
Ozone is a heavier gas than air, after switching off the circulating fan of the steriliser, STERYLIS tends to settle in
the recesses of the ground and migrate by the floor.
It is not recommended to sterilise rooms below ground level without mechanical ventilation.
It is forbidden to carry out the process of ozonisation by persons with olfactory disorders.
This equipment may only be used by children at least 16 years of age and by persons with reduced physical and
mental capabilities and those without experience and knowledge of the appliance, if supervision or instruction
regarding the use of the equipment is provided in a safe manner, so that the hazards are understandable. Children
should not play with the equipment. Children without supervision should not clean or maintain the equipment.
Moreover, the device should not be operated in places directly exposed to sunlight, with high humidity, in places where
chemicals are used and in rooms with very small cubic capacity. For the correct operation of the ozone sensors the device
must be used in an environment with the following parameters:
Temperature: -20 to 50˚C
Pressure: 900 to 1100 hPa
Do not clean the room intended for ozone sterilisation with chlorine-based cleaning agents or other chlorinated
substances. This can cause erroneous indications of the ozone sensor in the device and expose the user to danger.
The device is equipped with a pre-filter and an active carbon filter. In addition, there is one additional pre-filter in the
package, which must be used interchangeably with the activated carbon filter when starting the sterilisation mode.
The manufacturer does not guarantee that the ozone sterilisation process (carried out in an appropriate manner) will not