Sterylis VS Manual
| 25
To activate the front panel locking mode:
During active standby, silent disinfection, standard disinfection or filtration, press and hold the + and - buttons in the O
3
STERILIZATION section for at least 3 seconds.
To deactivate the front panel locking mode:
While the front panel lock mode is active, press and hold the + and - buttons in the O
3
STERILIZATION section for at least 3
seconds.
6.6 Changing the configuration parameters
To change the configuration parameters, press and hold the SET button in the OZONING TIME section in the standby
mode. Then enter the appropriate code to change the parameters.
Entering the access code:
The 4-digit code entered is shown on the display under O
3
STERILIZATION. The + and - buttons change the value by 1
active digit. An active number is the one that flashes at a high frequency, inactive numbers light up constantly. The
PROG.SET button changes the active digit to the next one on the right. After entering the correct code, confirm it by
pressing the yellow START button longer. If the code is incorrectly entered, the code XXXX or HHHH will be displayed and a
short beep will occur. In this case, the code must be re-entered.
You can change the following parameters:
Parameter
code
Description
Min value
Max value
Default value
C101
Operating concentration in AUTO mode [ppm]
1
10
7
C130
Siren operation in sterilisation mode [ON/OFF]
0
1
1
C138
50% reduction in the power of ozone generators
0
1
0
C150
Date setting - day of the month
1
31
15
C151
Date setting - month
1
12
7
C152
Date setting - year
2020
2120
2020
C200
FIRMWARE revision (read only)
-
-
FW_REV
The individual parameters are changed using the + and - buttons in the O
3
STERILIZATION section. The code of the
selected parameter is displayed in the O
3
STERILIZATION section. The current value of the selected parameter is displayed
in the O
3
CONCENTRATION section.
To change the value of the parameter, operate the + and - buttons in the O
3
CONCENTRATION. The active digit is selected using the SET button in the same section of the user panel. To confirm the
value of the selected configuration parameter, press and hold the PROG.SET button. The selected value will be confirmed
by a short beep and all active displays will flash.
To return to the standby mode, press and hold the SET button in the OZONING TIME section.
6.7 Working time counters
The device is equipped with the operating time counters listed in the table below. The user has the possibility to read the
indications of these counters and reset them.
To view or reset the counters, press and hold the SET button in the OZONING TIME section while the unit is in standby
mode. Then enter the appropriate code to change the parameters.
Entering the access code:
The 4-digit code entered is shown on the display under O
3
STERILIZATION. The + and - buttons change the value by 1
active digit. An active number is the one that flashes at a high frequency, inactive numbers light up constantly. The
PROG.SET button changes the active digit to the next one on the right. After entering the correct code, confirm it by