Sterylis VS Manual
| 28
The display in the O
3
CONCENTRATION section shows: The value of the current ozone concentration when the
concentration exceeds 0.1 ppm / characters - -.- - during the heating phase of the ozone sensor when the
concentration exceeds 0.1 ppm and when the selected sterilization mode / In other cases the display is inactive
The display under OZONING TIME indicates the characters - -.- -
6.8.4 SET-FACTORY-SETTINGS function (resetting to factory settings)
The function consists in restoring all configuration settings to their default values, in accordance with the table in Chapter
6.6.
6.8.5 SET-TIME function and setting the current date
The function is used to set the current time on the controller. Selecting the SET-TIME function displays the currently set
time on the display in the OZONING TIME section.
To set the current time:
The + and - buttons change the value by 1 active digit. The selection of the active digit is made by pressing the OK button
in the OZONING TIME section. The time is saved immediately after changing any number, no additional approval is
required.
To set the current date, go to configuration parameters C150, C151 and C152 and correctly define and validate the values
of these parameters in accordance with Chapter 6.6.
6.8.6 OZONE-MONITOR function
The function consists in measuring the instantaneous ozone concentration and displaying the measured value on the
display in section O
3
CONCENTRATION.