A pre-filter that is pulled out of the destination causes a violation of the limiting switch. If a limiting switch is violated
during the sterilisation process, the process will be stopped and completed and the device will go into standby mode. In
case of violation of the limiting switch in filtration or disinfection mode, the process will be stopped at the current moment
and the unit will go into pause mode. When the limiting switches of the filter cartridge are no longer violated (inserting the
filter cartridge back), the unit will resume the filtration/disinfection process. Violation of the end point also results in a
corresponding error on the display in the O
3
STERILIZATION section. If the end point violation stops, the error is
automatically reset.
The safety limiting switches provide protection in case of an attempt to change the air filters during operation.
7.2.2 Active carbon filter replacement (F2)
Make sure that the device is switched off and disconnected from the power supply.
Follow the instructions below:
1.
Remove the air filter cartridge out by pulling
it towards you
2.
Remove the air filter from the cartridge
3.
Place a new air filter in the cartridge. Make
sure that the direction of the arrow on the
cartridge is the same as the direction of the
arrow on the air filter
4.
Insert the filter cartridge back into the
device and push it
5.
The filter has been properly replaced