Sterylis VS Manual
| 32
The active carbon filter that is pulled out of the destination causes a violation of the limiting switch. If a limiting switch is
violated during the sterilisation process, the process will be stopped and completed and the device will go into standby
mode. In case of violation of the limiting switch in filtration or disinfection mode, the process will be stopped at the current
moment and the unit will go into pause mode. When the limiting switches of the filter cartridge are no longer violated
(inserting the filter cartridge back), the unit will resume the filtration/disinfection process. Violation of the end point also
results in a corresponding error on the display in the O
3
STERILIZATION section. If the end point violation stops, the error is
automatically reset.
The safety limiting switches provide protection in case of an attempt to change the air filters during operation.
7.3 Replacing the UV-C lamps
Regular replacement of the UV-C radiators is essential for proper operation. Exceeding the service life of UV light bulbs
may result in their burning out. It should also be remembered that with time the maximum dose of lamp radiation
decreases.
The lifetime of UV bulbs is approx. 9000 hours. This is at the same time the time after which a message indicating that
this time has been exceeded will be displayed.
Note! The replacement of UV-C fluorescent tubes in STERYLIS sterilisers may only be carried out by the
manufacturer's service or another authorised service centre!
7.4 Replacing the ozone generators
Regular replacement of the UV-C lamps is essential for proper operation. The lifetime of the generators is approx. 9000
hours.
Note! The replacement of ozone generators in STERYLIS sterilisers may only be carried out by the
manufacturer's service or another authorised service centre.
7.5 Ozone sensor service
The ozone sensor is one of the most important elements of the device, which determines its proper functioning and
safety. Regular recalibration and replacement after the expiry date of the ozone sensor is essential for proper and safe
operation. Re-calibration should take place every 1,000 hours of operation of the device in sterilization mode, while
replacement is related to the expiry date after 5 years from the purchase of the device.
Messages for recalibration and expiry date of the ozone sensor are displayed by means of codes F004 and F005 (See:
Chapter 9) and the warning beacon. When the counters are reset after recalibration or sensor replacement (according to
the instructions in Chapter 6.5), the signaling by means of messages and the warning beacon is switched off.
Note! The replacement and re-calibration of the ozone sensor in STERYLIS sterilizers may only be carried
out by the manufacturer's service or another authorised service centre
.