The CliniMACS Formulation Unit may be used repeatedly. It is not intended for
disposal after single use. Contact the local authority governing electrical power
supply, building construction, maintenance, or safety for more information
regarding the installation of the equipment.
Risk of damage to the tubing set. The tubing set can be damaged if the tubing is
streched by the bags hanging over the edge of the table. Leave sufficient space
(approx. 15 cm) for bags for tubing set installation in front of the CliniMACS
Formulation Unit.
3.1.2 Hazards of electric shock and spread of fire
Electrical instruments pose the risk of an electric shock, electrical short, and
overheating. Electric shock may lead to severe personal injury or death.
An electrical short and overheating may lead to the spread of fire.
The CliniMACS Formulation Unit is a Protection Class I device and may only be
connected to the CliniMACS Prodigy. The CliniMACS Prodigy itself may only be
plugged into an outlet with a grounded connection.
Electronic equipment may emit sparks that could ignite combustible vapors or
dusts leading to explosion or spread of fire.
Risk of electric shock, electrical short, and spread of fire. Electric shock
may lead to severe personal injury or death. The housing of the
instrument reduces the risks of electric shock, electrical short, and spread
of fire. Opening any cover of the instrument interrupts this protective
measure. Do not remove or penetrate any cover of the instrument. Only
authorized local Miltenyi Biotec Service Provider may open the
instrument or exchange spare parts. Do not use the instrument in case of
visible damage or if it has been dropped. Contact
Miltenyi Biotec Technical Support.
Risk of electric shock or damage to the instrument. Risk of electric shock
or damage to the instrument if the instrument is cleaned with excessive
amount of cleaning or while switched on. Before cleaning or disinfection
of the CliniMACS Formulation Unit, disconnect the instrument from the
power supply. Unplug the mains plug of the CliniMACS Prodigy to
disconnect the CliniMACS Formulation Unit from the power supply. Only
use the originally supplied power cord.