30
4.5 Cleaning and disinfection
Risk of electric shock or damage to the instrument if the instrument is
cleaned with excessive amount of cleaning or while switched on. Only
clean the CliniMACS Formulation Unit when the CliniMACS Prodigy is
switched off and the power cord of the CliniMACS Prodigy is unplugged.
Avoid ingress of any liquid into the pinch valves.
Clean the instrument with one of the following disinfectants, which are
compatible with the surface of the instrument:
• Aldehyde up to 3%, e.g., Kohrsolin® FF 3% or
• Ethanol up to 80%
Safety and performance of the instrument may be compromised if other
than the above mentioned cleaning methods are used. UV may damage
plastic parts of the instrument. Do not use UV irradiation. The instrument
should not be sterilized, e.g., with H
2
O
2
.
Dust off the valves using a paper towel.
IMPORTANT
The surface of the CliniMACS Formulation Unit should be cleaned at regular intervals
and after each application. Cleaning after unpacking before installation is also
recommended.